Processing step
Items Needed
Machine cleaning/
Erbe irrigation adapter 20132-087
disinfection
9.5
Precleaning
Preclean the product up to a maximum of 2 hours after use.
For precleaning, you will need water, a 3 % hydrogen peroxide
solution, as well as a mildly alkaline (pH value between 7 and 10)
cleaning agent suitable for ultrasonic use with medical devices
comprised of plastic and metal.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Immerse the product completely in a hydrogen peroxide
bath, ensuring that the product does not come into contact
with other components in the bath (immersion time: at least
5 minutes).
3. Screw the irrigation adapter (Art. No. 20132-087) onto the
APC applicator.
4. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
5. If available: Retract and extend the movable electrode sever-
al times. The electrode must be retracted for the following
processing steps.
6. Remove the syringe. Leave the irrigation adapter on the
product for the next processing steps.
7. Immerse the product completely in an ultrasonic cleaning
bath. Do not allow the product to come into contact with oth-
er components in the bath (immersion time in accordance
with the instructions provided by the manufacturer of the
cleaning agent, ultrasonic cleaning: 10 minutes).