Section 3
Electrical Service Requirements (Standard temperature chillers):
ThermoFlex3500/5000
ThermoFlex7500/10000
ThermoFlex7500/10000
MCA = Minimum Current Ampacity
MOPD = Maximum Overcurrent Protective Device
Values reflect those on the nameplate located on the rear of the chiller.
3-4
ThermoFlex
Voltage ±10%
200 VAC P 1, P 2 Pump
200 VAC T 1, P 3 Pump
200 VAC P 4 Pump
208-230 VAC P 1, P 2 Pump
208-230 VAC T 1, P 3 Pump
208-230 VAC P 4 Pump
230 VAC P 1 - P 4 Pump
Voltage ±10%
(Air-cooled)
200 VAC P 2 Pump
200 VAC P 3 Pump
200 VAC P 5 Pump
200 VAC T 5 Pump
208-230 VAC P 2 Pump
208-230 VAC P 3 Pump
208-230 VAC P 5 Pump
208-230 VAC T 5 Pump
400 VAC P 2 Pump
400 VAC P 3 Pump
400 VAC P 5 Pump
400 VAC T 5 Pump
Voltage ±10%
(Water-cooled)
200 VAC P 2 Pump
200 VAC P 3 Pump
200 VAC P 5 Pump
200 VAC T 5 Pump
208-230 VAC P 2 Pump
208-230 VAC P 3 Pump
208-230 VAC P 5 Pump
208-230 VAC T 5 Pump
400 VAC P 2 Pump
400 VAC P 3 Pump
400 VAC P 5 Pump
400 VAC T 5 Pump
Frequency
Phase
50 Hz
1Ø
50 Hz
1Ø
50 Hz
1Ø
60 Hz
1Ø
60 Hz
1Ø
60 Hz
1Ø
50 Hz
1Ø
Frequency
Phase
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
Frequency
Phase
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
Branch Circuit
Line Cord
Requirements
Plug
15A
6-15P
20A
6-20P
30A
L6-30P
15A
6-15P
20A
6-20P
30A
L6-30P
*16A
, 15A
, 13A
-
1
2
3
MCA
MOPD
Line Cord
Plug
16.5
30
L15-20P
18.7
30
L15-30P
22.3
35
L15-30P
17.3
30
L15-20P
16.5
30
L15-20P
18.7
30
L15-30P
22.3
35
L15-30P
17.3
30
L15-20P
10.9
20
IEC309
9.6
15
IEC309
11.8
15
IEC309
8.7
15
IEC309
MCA
MOPD
Line Cord
Plug
16.2
30
L15-20P
18.4
30
L15-30P
22.0
35
L15-30P
17.0
30
L15-20P
16.2
30
L15-20P
18.4
30
L15-30P
22.0
35
L15-30P
17.0
30
L15-20P
10.6
20
IEC309
9.3
15
IEC309
11.5
20
IEC309
8.4
15
IEC309
Continued on next page.
Thermo Scientific