Electrical Service Requirements (High temperature chillers):
ThermoFlex3500/5000
ThermoFlex7500/10000
ThermoFlex7500/10000
MCA = Minimum Current Ampacity
MOPD = Maximum Overcurrent Protective Device
Values reflect those on the nameplate located on the rear of the chiller.
Thermo Scientific
Voltage ±10%
200 VAC P 1 P 2 Pump
200 VAC T 1 P 3 P 4 Pump
208-230 VAC P 1 P 2 Pump
208-230 VAC T 1 P 3 P 4 Pump
230 VAC P 1- P 4, T1 Pump
Voltage ±10%
(Air-cooled)
200 VAC P 2 Pump
200 VAC P 3 Pump
200 VAC P 5 Pump
200 VAC T 5 Pump
208-230 VAC P 2 Pump
208-230 VAC P 3 Pump
208-230 VAC P 5 Pump
208-230 VAC T 5 Pump
400 VAC P 2 Pump
400 VAC P 3 Pump
400 VAC P 5 Pump
400 VAC T 5 Pump
Voltage ±10%
(Water-cooled)
200 VAC P 2 Pump
200 VAC P 3 Pump
200 VAC P 5 Pump
200 VAC T 5 Pump
208-230 VAC P 2 Pump
208-230 VAC P 3 Pump
208-230 VAC P 5 Pump
208-230 VAC T 5 Pump
400 VAC P 2 Pump
400 VAC P 3 Pump
400 VAC P 5Pump
400 VAC T 5 Pump
Frequency
Phase
50 Hz
1Ø
50 Hz
1Ø
60 Hz
1Ø
60 Hz
1Ø
50 Hz
1Ø
Frequency
Phase
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
Frequency
Phase
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
60 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
50 Hz
3Ø
Section 3
Branch Circuit
Line Cord
Requirements
Plug
30A
6-30P
40A
Hard wire
30A
6-30P
40A
Hard wire
32A
IEC309
MCA
MOPD
Line Cord
Plug
35.7
45
Hard wire
37.9
45
Hard wire
41.5
50
Hard wire
36.4
45
Hard wire
35.7
45
Hard wire
37.9
45
Hard wire
41.5
50
Hard wire
36.4
45
Hard wire
19.2
25
Hard wire
17.9
25
Hard wire
20.1
25
Hard wire
17.0
25
Hard wire
MCA
MOPD
Line Cord
Plug
35.3
45
L15-20P
37.5
45
L15-20P
41.1
50
L15-30P
36.1
45
L15-30P
35.3
45
L15-20P
37.5
45
L15-20P
41.1
50
L15-30P
361
45
L15-30P
18.8
25
IEC309
17.5
25
IEC309
19.7
25
IEC309
16.7
25
IEC309
Continued on next page.
ThermoFlex
3-7