Application
Figure 3-4 Typical Plumbing
Connections (2 of 2)
Water-cooled Chillers
For water-cooled chillers the facility water plumbing connections are also located
on the rear and are labeled
ouTLET. The connections are ½" Female NPT for ThermoFlex900 -
5000, ¾" Female NPT for ThermoFlex7500 - 10000. Both connections for
ThermoFlex900 to 10000 are cast bronze.
Connect the
FACILITY InLeT to your facility water supply. Connect
the
FACILITY ouTLeT to your facility water return or drain. Ensure
all connections are secure and that the proper sealant/lubricant for the fitting
material is used. (If Teflon
first thread as it could shred and get into the fluid.)
Figure 3-5 Typical Plumbing Connections, Water-cooled Chillers
Connect the PRoCESS ouTLET
the fluid inlet on your application. Connect the
PRoCeSS InLeT
your application. Ensure all connections are
secure and that the proper sealant/lubricant
for the fitting material is used. (If Teflon
is used, ensure the tape does not overhang the
first thread as it could shred and get into the
fluid.) Keep the distance between the chiller
and the instrument being cooled as short as
possible. Ensure tubing is straight and without
bends. If diameter reductions are required,
make them at the inlet and outlet of your
application, not at the ThermoFlex.
FACILITY InLeT and
tape is used, ensure the tape does not overhang the
®
Section 3
to
to the fluid outlet on
tape
®
FACILITY
See Section 2 for the specific
locations on your chiller.
ThermoFlex
3-11