Section 4
RESISTIVITY (Optional)
Note
his option is only available on standard-temp chillers.
T
RESISTIVITY enables/configures the resistivity option. With DE-I CONTROL highlighted press
enter to enable the feature. Next set the limits to the desired values.
RESISTIVITY
DE-I CONTROL
TEMP. COMPENSATION
With TEMP. COMPENSATION highlighted press enter to turn compensation off or on.
RESISTIVITY
DE-I CONTROL
TEMP. COMPENSATION
When the desired changes are made press the left arrow, or esc button, to return to the SETTINGS
display.
RESTORE DEFAULTS
RESTORE DEFAULTS restores the controller back to factory default values. A reset is automatically
performed whenever new firmware is installed.
RESTORE
DEFAULT SETTINGS?
YES
NO
4-20 ThermoFlex
De I Control
Setpoint
Interval
Warn High
Warn Low
Resistivity
Setpoint:
Interval
Warning High:
Warning Low:
TEMPERATURE
COMPENSATION
None
NaCl- Water
Press the left arrow, or esc button, to return to the
SETTINGS display.
x.x MΩ-cm
x.x MΩ-cm
x.x MΩ-cm
x.x MΩ-cm
Range
Default
0.2 to 3.0
1.0 MΩ-cm
0.1 to 0.5
0.1 MΩ-cm
0.0 to 3.5
3.0 MΩ-cm
0.0 to 3.5
0.5 MΩ-cm
Thermo Scientific