Thermo Scientific ThermoFlex Manual De Instrucciones página 85

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 26
Section 2
Pumping Capacity
Turbine Pump T 0
2-8
ThermoFlex
Bar
PSID
160
10
140
120
8
100
6
80
60
4
40
2
20
2.5
0.5
• Pump curves are nominal values. Pressure values for turbine pumps are differential pressures
between the inlet and the outlet of the chiller.
• Pump performance results were obtained with no restrictions on the return to the system or
with any options installed. For example, utilizing the DI option will result in a 0.5 gpm flow
reduction .
• Specifications obtained at sea level using water as the recirculating fluid, at a 20°C process
setpoint, 25°C ambient condition, at nominal operating voltage. Other fluids, fluid temperatures,
ambient temperatures, altitude or operating voltages will affect performance. See Section 3.
• Thermo Fisher Scientific reserves the right to change specifications without notice.
60 Hz
50 Hz
5.0
7.5
1.0
1.5
2.0
Flow Rate
10.0
12.5 LPM
2.5
3.0
GPM
Thermo Scientific