Facility Water
Requirements
(water-cooled chillers)
CAUTION
ThermoFlex1400
ThermoFlex2500
Thermo Scientific
Facility Water Maximum Inlet Pressure must not exceed 150 PSIG.
Facility Water Maximum Pressure Differential must not exceed 50 PSID.
(Pressure Differential = Inlet Pressure - Outlet Pressure)
Contact us before using facility water that is above 35°C.
The facility water must meet the following conditions for the chillers to
maintain their full rated capacity.