FLUID LEVEL
FLUID LEVEL is used to view/change the chiller's reservoir level fault and warning settings. If the
chiller exceeds the fault setting it will shut down, the controller will display a fault message and, if
enabled, sound the alarm. If the chiller exceeds the warning setting the chiller will continue to run, the
controller will display a warning message and, if enabled, sound the alarm.
If a change to the setting is needed, press the arrows to highlight the desired line and then press the
enter button. The highlight will flash. Press the up or down arrow to change the value. Once the
desired setting is displayed press enter to accept the change and stop the flashing.
Note You cannot set the Low Warning value lower than the Low Fault value.
When the desired changes are made press the left arrow, or esc button, to return to the SETTINGS
display.
FLUID LEVEL
Low Warning
Low Fault
AUTO REFILL (Optional)
AUTO REFILL is used to view/change the chiller's optional auto refill settings. On is the % of fluid
level in the reservoir needed to turn the option on. Time Out is the maximum time the option will
operate. Setting the time to 0 disables the option.
When the desired changes are made press the left arrow, or esc button, to return to the SETTINGS
display.
AUTO REFILL
On
Time Out
Thermo Scientific
Low Warning/Fault Range/Default:
Heater dependent - see below
x%
Heater
x%
None
1 kW:
2.3 kW:
5.0 kW:
4.6 kW:
Heater
None
1 kW:
2.3 kW:
5.0 kW:
4.6 kW:
On % Default: Heater dependent - see below
Heater
None
x%
1 kW:
xxsec
2.3 kW:
5.0 kW:
4.6 kW:
Time Out Range: 0 - 900 seconds
Time Out Default: 45 seconds
80 seconds for ThermoFlex7500/10000
Setting the time to 0 disables the option.
Warning Range
Default
6 - 100%
6%
58 - 100%
58%
58 or 69 - 100%
58 or 69%
87 - 100%
87%
87 - 100%
87%
Fault Range
Default
0 - 100%
0%
52 - 100%
52%
63 or 52 - 100%
63 or 52%
81 - 100%
81%
81 - 100%
81%
Range
Default
70 - 100%
70%
70 - 100%
70%
69 or 70 - 100%
69 or 70%
87 - 100%
87%
87 - 100%
87%