EN - ENGLISH
APPENDIX B: DATA DOWNLOAD FILE FORMAT
The ICP Monitor can download data to an external USB flash
drive with the following specifications:
• USB 2.0
• 8 to 32 GB capacity
• FAT32 format
Data available for download include:
• Last 1 hour or 24 hours of ICP Waveform data (100 samples
per second)
• Trend data, including last 24 hours or all 14 days of:
• Mean ICP values (1 sample every 10 seconds)
• Physiological alarm conditions
• Event markers
• Last or all available screen shots (50 maximum)
Data is stored in the following format:
• Waveform data - Tab delimited text file with 3 columns:
Date
Time
YYYY/MM/DD
HH:mm:ss
• Trend data - Tab delimited text file with 5 columns:
Date
Time
YYYY/MM/DD HH:mm:ss XX.X
• Images – Bitmap (BMP) file
The data will be stored in the USB flash drive in files following a
specific file naming convention:
Waveform data:
• CereLink-wave-SNxxxxxxx-YYYYMMDD-HHmm(#)
Trend data:
• CereLink-trend-SNxxxxxxx-YYYYMMDD-HHmm
Screen shots:
• CereLink-IMG-SNxxxxxxx-YYYYMMDD-HHmm
Where:
• YYYY = Year
• MM = Month
• DD = Day
• HH = Hour (24 hour format)
• mm = minutes
• (#) = file number
• xxxxxxx = Sensor serial number (uses "x" if no sensor is
connected)
ICP [mmHg]
XX.X
ICP [mmHg] Alarm[H/L] Marker[1-99]
X
X
File Name Examples:
• Monitoring started 12/01/2017 at 3:45 PM
• Sensor Serial number = 1234567
• Download initiated to USB flash drive at 12/6/2017
3:54 PM
• Patient has been monitored for 5 days and 9 minutes.
Scenario 1: Last 24 hours of trend data
• CereLink-trend-SN1234567-20171206-1554
Scenario 2: All available trend data
• CereLink-trend-SN1234567-20171206-1554
Note: Scenarios 1 and 2 will create the same file name, but
with different amount of data
Scenario 3: Last 1 hour waveform
• CereLink-wave-SN1234567-20171206-1554(1)
Scenario 4: Last 24 hours waveform
• CereLink-wave-SN1234567-20171206-1554(1)
• CereLink-wave-SN1234567-20171206-1554(2)
• CereLink-wave-SN1234567-20171206-1554(3)
:
:
• CereLink-wave-SN1234567-20171206-1554(23)
• CereLink-wave-SN1234567-20171206-1554(24)
Note: Scenarios 3 and 4 will create the same file name for
the first hour data, containing the same information
Scenario 5: 2 images in memory, one taken before
sensor was connected
• CereLink-IMG-SN1234567-20171205-1146
• CereLink-IMG-SNxxxxxxx-20171201-1501