EN - ENGLISH
MAINTENANCE
Preventive Maintenance
While this device does not require periodic repair or calibration, general care, handling and maintenance will extend the life of the
device. Such practices the user must observe include:
• The integrity of equipment, plastic covers, LCD and interconnecting cables is sound.
• External equipment markings remain legible.
• The equipment is free of dirt or debris likely to impair safety or performance (follow the instructions for wiping the device).
• The controls and display function in accordance with the Instruction Manual.
• The rechargeable battery is charged.
• After exposure to extreme temperatures allow the equipment to return to room temperature before use. Do not allow the ICP
Monitor to stay in extreme temperatures for prolonged periods of time.
• Do not drop the device.
Cerelink ICP Monitor operating efficiency is directly related to the physical condition of the unit. If any damage or abuse is observed
the device must be returned to the nearest repair facility (see below locations).
The device has double insulation and does not require a safety connection to electrical earth (ground). Leakage current, dielectric and
ground bond testing are not required as the system uses an approved Class II external Power Adapter (transformer).
Annual Preventive Maintenance and Testing conducted by Integra LifeSciences is recommended to inspect for any possible wear
that will degrade your CereLink ICP Monitor System's essential performance or operation, including Electrical Safety Testing per
62353:2014.
Being proactive in your CereLink ICP Monitor Maintenance program ensures your equipment will run at optimal capability for longer,
allowing you to get the most from your monitor, protecting your patients, users and your financial investment.
Perform the periodic maintenance procedures according to the schedule provided below:
Task
Wiping Exterior Surfaces
Preventative Maintenance
Replace the Battery
Service and Repair
Send the ICP Monitor for service or repair to the nearest repair facility:
US Service Center
Integra Neurosciences
5965 Pacific Center Blvd
Suite 705
San Diego, CA 92121
Tel: 800-815-1115 Option 5
Fax: 858-455-5874
Product End of Life
The monitor contains electrical components. Dispose of the equipment in accordance with local ordinances.
PRODUCT INFORMATION DISCLOSURE
INTEGRA LIFESCIENCES CORPORATION ("INTEGRA") HAS EXERCISED REASONABLE CARE IN THE SELECTION OF MATERIALS
AND THE MANUFACTURE OF THESE PRODUCTS. INTEGRA WARRANTS THAT THESE PRODUCTS SHALL CONFORM TO THE
PRODUCT LIMITED WARRANTY AS PROVIDED IN THE PRODUCT LABELING OR APPLICABLE PRODUCT CATALOG. THIS WARRANTY
IS EXCLUSIVE AND INTEGRA DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT
LIMITED TO, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. INTEGRA SHALL NOT
BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL LOSS, DAMAGE, OR EXPENSE, DIRECTLY OR INDIRECTLY ARISING FROM USE
OF THESE PRODUCTS. INTEGRA NEITHER ASSUMES NOR AUTHORIZES ANY PERSON TO ASSUME ANY OTHER OR ADDITIONAL
LIABILITY OR RESPONSIBILITY IN CONNECTION WITH THESE PRODUCTS.
Frequency
Daily, when in use
Annually
Every 3 years or when battery performance
has deteriorated
Europe, Middle East and Africa Service Center
Integra Nuerosciences GmbH
Halskestrasse 9
Ratingen 40880
Germany
Tel: +49 2102 5535 6150
Fax: +49 2102 942 4872
Procedure
Refer to WIPING, STORAGE & TROUBLESHOOTING
section
Refer to SERVICE & REPAIR section
Refer to REPLACING THE BATTERY section
Asia Pacific Service Center
Integra Nuerosciences Pty . Ltd .
Unit 3, 24-30 Winterton Road
CLAYTON, VIC. 3168, Australia
Tel: +613 85400400
Fax: +613 95400004
32