Technical Specifications - Integra Codman CereLink Manual Del Usuario

Ocultar thumbs Ver también para Codman CereLink:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 153
EN - ENGLISH

TECHNICAL SPECIFICATIONS

SENSING TECHNOLOGY – CereLink ICP Sensor
Pressure Sensor Technology
Configurations
Implant Duration
Zeroing
Zeroing medium
Reconnection to other CereLink monitors
MRI
Pressure Accuracy
Pressure drift
Temperature coefficient
Diameter
INPUT PRESSURE CHANNEL
Operating Range
Sensor Excitation Voltage
Pressure Signal Sensitivity
Input Impedance
Pressure Signal Band Width
Zero Range
Input Channel Accuracy
Input Channel Resolution
DISPLAY
Type
Size
Viewing Angle
ALARMS
Alarm Parameter
High Alarm Range
Low Alarm Range
High/Low Alarm Step Resolution
High Alarm Default Value
Low Alarm Default Value
Strain gauge
Basic Kit
Metal Bolt
Plastic Bolt
Ventricular Catheter
< 30 days
Automatic, stored in sensor memory
Sterile water or sterile saline
YES
1.5T and 3T
a) ±2mmHg over the range
-20mmHg to 20mmHg
b) ±10% of reading over the range
-50 to -20 and 20 to 150mmHg
< ±5mmHg over 30 days
±2mmHg when going from 23°C to 38°C
Maximum tip diameter of 1.295mm
Ventricular Catheter: 3.5mm
-50 to 150mmHg
5VDC +/- 0.01V
5uV/Vex/mmHg
10 MΩ nominal
0 to 25Hz (-3dB)
+/- 50 mmHg
+/- 0.5mmHg over the range -50 to 50mmHg and
+/-1% over the rage 50 to 150mmHg (exclusive of sensor)
0.1mmHg
TFT
1024 (H) x (RGB) x 600 (V)
LCD with projected capacitive touchscreen
7.0"
+/- 30o
Mean ICP
-50 to 150mmHg
-50 to 150mmHg
1 mmHg
20mmHg
0 mmHg
28

Capítulos

Tabla de contenido
loading

Tabla de contenido