Integra Codman CereLink Manual Del Usuario página 34

Ocultar thumbs Ver también para Codman CereLink:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 153
EN - ENGLISH
APPENDIX A: ELECTROMAGNETIC COMPATIBILITY – GUIDANCE AND MANUFACTURER'S
DECLARATIONS
TABLE 3: Guidance and Manufacturer's Declaration – Electromagnetic Immunity .
The ICP Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the ICP
Monitor should ensure that it is used in such an environment.
affect its life expectancy.
IMMUNITY test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
80% AM at 1 kHz
RF Proximity
Band
Fields
380 - 390 MHz
IEC 61000-4-3
430 - 470 MHz
704 - 787 MHz
800 - 960 MHz
1,700 - 1,990 MHz
2,400 - 2,570 MHz
5,100 - 5,800 MHz
Deviations from emission and immunity environment for this device may
Compliance level
Electromagnetic environment – guidance
V1 = 3Vrms
Portable and mobile communications equipment should be
seperated from the ICP Monitor by no less than the distances
calculated /listed below:
E1 = 3V/m
D=(3.5/V1)(√ P) 150 kHz to 80 MHz
D=(3.5/E1)(√ P) 80 to 800 MHz
D=(7/E1)(√ P) 800 MHz to 2.5 GHz
where P is the max power in watts and D is the recommended
Immunity Test Level
seperation distance in meters.
27 V/m
28 V/m
Field strengths from fixed transmitters, as determined by
9 V/m
an electromagnetic site survey, should be less than the
28 V/m
compliance levels (V1 and E1)
28 V/m
28 V/m
Interference may occur in the vicinity of equipment containing
9 V/m
a transmitter.
34

Capítulos

Tabla de contenido
loading

Tabla de contenido