Battery and Charging Safety

Battery
• Do not attempt to replace the lithium-ion battery.
• Do not attempt to replace the lithium-ion battery
with non-approved batteries. Incorrect replace-
ment of the battery could result in damage to
EXOGEN. The battery should only be serviced by
Bioventus trained personnel.
• Be sure to use only the USB battery charger (Part
Number 71034460) provided with the system.
Other battery chargers may cause battery
overheating and damage the battery, EXOGEN,
the battery chargers or the user.
• Do not use an extension cord with the battery
charger as this may cause overheating.
• Do not use the battery charger with other devices
as this may damage the battery charger and/or the
other device.
• If the battery area on EXOGEN or the battery
charger becomes excessively warm, discontinue
using and contact Customer Service at 1-800-
836-4080.
Labeling LBL-000319 [D] RELEASED
Charging
• Charge the battery to at least 25% capacity (one
bar) before attempting to perform a treatment
when the battery is used for the first time or after
prolonged storage.
• The battery will charge whether
EXOGEN is turned off or on.
• If the battery power decreases
quickly even after recharging for
many hours contact Customer
Service at 1-800-836-4080.
Do not recharge the battery in any of the
following locations;
• Where the ambient temperature is below 0ºC or
above 45ºC
• Damp or wet location and/or near water
• Outside (use indoors only)
• Within the reach of small children
• With the battery charger cable stretched across a
floor or other areas where people walk that would
cause a tripping hazard
• On floor or other area where EXOGEN or the cable
may be damaged by people walking on them