Setting Up The Unit - Dentsply Sirona Calamus Dual A1300 Manual Del Usuario

Ocultar thumbs Ver también para Calamus Dual A1300:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 163
STEP-BY-STEP INSTRUCTIONS
6.2

SETTING UP THE UNIT

1. Unpack the console and check that the Voltage Selector is set to the
proper voltage. Use the 115V position for 110-120V 60Hz voltages,
and the 230V position for 220-250V 50Hz voltages. Always unplug the
unit before changing fuses or adjusting the voltage selector. To change
voltage, use a flat head screwdriver to turn the Voltage Selector on the
back of the console to the proper voltage setting. To change voltage,
replace the fuse to match voltage
36). Replace with 0,3A, 230/250V slow blow fuse (0,6A for 115V).
For replacement, press on the black plastic clip of the fuse, pull out
and replace with the correct fuse.
2. Remove packaging from hand pieces. Clean the exterior of the hand
pieces with a soft cloth moistened with mild non-chlorinated detergent
or disinfecting solution. Do not use disinfecting solutions which contain
phenol, anticorrosives, very acidic or very alkaline solutions - do not
submerge hand pieces.
Place the hand pieces into their respective holders - Pack on the
left side and Flow on the right side.
The forward portion of the Flow hand piece becomes warm during use.
The heat shield (included with the system) may be used optionally to
reduce the Flow hand piece surface temperature. If the heat shield is
not utilized, avoid contact with the forward portion of the hand piece.
Sterilize the heat shield before first use and between each patient use.
See "DISINFECTION, CLEANING AND STERILISATION", page 25.
3. Connect hand piece cables to console: The hand piece cables are
color-coded and keyed to only fit the proper receptacle on the console.
With the arrow on the connector facing up, gently push the connector
into the console receptacle.
4. Attach the power cord to the back of the console and plug into a
grounded electrical outlet.
5. Sterilize Pluggers and Thermal Response Tip: Steam autoclave for
18 minutes at 134°C.
10/40
(See "SPECIFICATIONS", page
See Fig.
2.
BENCALDDFUWEB / Rev.04 / 02-2019 (Old ZF 1902103.EN)
Tabla de contenido

Solución de problemas

loading

Tabla de contenido