ENGLISH
8.3
Characteristics of the water to be used
TABLE SHOWING THE QUALITY LEVELS LAID BY THE UNI EN 13060
Evaporation residue
Silicon oxide
Iron
Cadmium
Lead
mg/l Remains of heavy metals apart from iron,
cardium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
Ph value (acidity level)
Appearance
Hardness (and alkaline earth ions)
8.4
Loading the material into the autoclave
Place the materials to be sterilized on the trays provided:
never superpose the materials
never bring the materials into contact with the sterilization chamber or the closing door
put the scissors and dental forceps with the blade open
When you have placed all the instruments close the door and you can read DOOR CLOSED
on the display.
18
CEN STANDARD UNI EN 13060
FIG.24
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5 7,5
Transparent, clear, without deposits
0.02
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
s/cm
mmol/l