Messages of error are emphasized through an Alfa-numerical code, consisting in a letter and
three numbers.
The "E" code is relative to errors of the operator, which can be corrected by same operator.
The "A" code is relative to alarms, irregularities of the autoclave, after having carried out the
remedy, if the problem persists it is necessary to request telephonic assistance.
If an alarm message is displayed (CODE "A") the cycle is to be considered
aborted: it will be necessary to repeat all the preparation and sterilization
operations.
To reset alarms and errors, maintain pressed the buttons here indicated until the intervention
of a
BEEP"
"
B-DOOR
 
 
ERROR
E 200
Low water level
Low level water cleaned up with loadingfrom
E 250
demineralizer
E 300
Maximum level used water
E 401
Door open
E 451
Door unlocked
A 250
Cycle begins without water
A 001
Interrupted cycle
Pressure anomaly after loading water into
A 003
the room
A 011
Irregular pc display
Hatch opende during the cycle or problem
A 401
closing
A 501
Pressure not maintened
A 551
Irregular 1° temperature sensor
24
 
MESSAGES OF ERROR OR ALARMS
+
B-PUMP
CAUSE
+
B-SELECT
REMEDY
 Fill tank
 Fill tank
 Empty tank
 Colse door with attention
 Open door and close again
 Control demineralizer device
 Reset and restart cycle
 Reset, adjust the door (see
the maintenance procedure)
 Shut off and restart autoclave
 Check the closing system
 Reset and restart cycle
 Reset and restart cycle
 