8.3 – MANUAL LOADING OF PURE WATER TANK
In the case the water loading pump does not function, filling the tank can be done manually by the operator
using the following method:
-turn on autoclave
-twist off the top located on the front of the autoclave (fig. 5 pos.1)
-insert in this opening the hose connected to the funnel
-pour distilled water in the funnel, keeping it higher than the loading top
-add until the led of maximum level turns on (fig.6 pos.1)
-loading completed, twist on the top
FIGURE 5
Frontal view 1.2
pos. 1
8.4 – CHARACTERISTICS OF WATER TO USE
Attachments
TABLE OF QUALITATIVE LEVELS ESTABLISHED BY REGULATION DIN EN 285
Evaporation residual
Silicon oxidation
Iron
Cadmium
Lead
Residual of heavy metals such as iron,
cadmium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
PH value (grade of acidity)
Appearance
Hardness (E alcaline-earth ions)
CEN STANDARD DIN EN 285
≤
≤
≤
≤
≤
≤
≤
≤
≤
Colorless, limpid, without deposits
≤
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5 to 7
0.02
53
FIGURE 6
Frontal view 1.1
pos. 1
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
µs/cm
mmol/l