ENGLISH
The following steps are recommended by the manufacturer. These operations must be
performed only by trained technicians authorized by the manufacturer
Clean the chamber
Clean the metal filter in the chamber
Replace the seal of the chamber door (DANA038)
Lubricate the closing system of the chamber door
Adjust the closing system of the chamber door
Replace the bacteriological filter (DAVA101)
Replace the water loading filter (DARA054)
Clean the Y brass filter downstream of the radiator
Clean the solenoid valves or replace it if necessary
EVERY ONE
Clean the tanks
YEAR /
Check the tank level probes
EVERY 800
Clean the radiator and cooling fans
CYCLES
Check the condition of the electrical/pneumatic circuit
Check that there are no traces of disinfectant in the pneumatic circuit.
Please note: ignoring traces of disinfectant may result in the need for
supplementary maintenance of the equipment
Check the safety valve
Perform validation in compliance with the local standards and on the basis
defined by the local standards
Perform the electrical safety tests in compliance with the local standards
and on the basis defined by the local standards
Measure the conductivity of water
AFTER 1600
Check the heating part
CYCLES
Replace the N/C solenoid valve (CEECG021)
AFTER 2400
CYCLES
Replace the heating part (DXBA835 or DHYA035)
A4
.
...................... µS/cm