4.1 Digestion process operation
The suggested sequence for preparing and starting the complete digestion
system is:
• Prepare the reagents and the samples in the digestion tubes. See the digestor's
instructions.
• Program the temperature controller for a three steps digestion cycle. According
to the digestor instructions manual.
• Insert the sample tubes in the block-digestor and cover them with the fume
collector.
• Check that the vacuum pump is ready. That is, the water tank and the external
circulating in OFF.
• Check that the scrubber bottles have the sodium carbonate solution.
• Assembly the hoses:
• Start the digestion cycle from the temperature controller.
• Start the vacuum pump.
• Visually check every 30 or 40 minutes, the digestion process.
• When finishing the digestion process, the temperature process beeps.
Then carry out the following operations, with caution:
• Remove the tubes rack from the digester without disconnecting any
hose and leave it on the holder.
• Keep the vacuum pump on.
• Remove the fume collector from the tube rack and leave it on the
holder.
• Fume collector cleaning:
You can simply clean the fume collector with water.
• Water tank emptying:
To empty the tank, just remove the «emptying tube» from the upper connector.
• Water circulating function:
The vacuum pump can also work as a water closed circuit.
- Connect the hose to the «water circulating out port».
- Place the «water circulating switch» in the ON position.
- When starting the pump, this will move the tank water to the «water circulating
out port».