Technical Specification (Digestor Block); Installation - Selecta DQO 6 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES 80158 REV D
Block:
Tubes:
Electric power: (W)
Weight (kg)
Process control:
Temperature range:
Load maximum power:
Temperature stability:
Memory:
Maximim time per step:
RS-232 serial communication:
"Scrubber" unit 4001611
Process controller «RAT»
J
.P. SELECTA s.a.u.
3.2
6x MACRO
12x MACRO
6x Ø42
12x Ø42
1500
2100
18
25
45..450ºC
10A
±1ºC
20 programs of 4 steps
600 minutes
Optional
4.
Installing a Kjeldahl digestion system requires a suitable place for it specifically,
due to acid fumes generated during digestion.
The installation within cab¡ns, hoods, ... should consider the compatibility of the
materials used with acidic fumes.
It is best to neutralize the acid fumes through a "scrubber" and a suction pump.
The scrubber that contains a concentrated solution of calcium carbonate, the
effect of the vacuum produced by the "Pump" digestion gases pass through the
calcium carbonate solution that neutralizes the gases released by the evaporation
of sulfuric acid digestion.
Fume collector
Digestion block
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
OCT/2019
(Sujetas a modificaciones sin previo aviso)
Technical specification (Digestor blocks)
20x MACRO
12x MICRO
20x Ø42
12x Ø42
2500
1500
31
16

Installation

4001611
"Scrubber" unit
24x MICRO
40x MICRO
24x Ø42
40x Ø42
2100
2500
22
27
4001612
Water recirculating vacuum
pump
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Pag.: 22

Hide quick links:

Tabla de contenido
loading