719 S Titrino
ON means that the Titrino reports automatically the moment the correspond-
ing change occurs.
.Status:
Global switch for all set AutoInfo.
.P
PowerOn: Simulation of power on (3.2.2.87). Not from mains.
Messages from node .T, Titrator:
.T.R
Ready: Status 'Ready' has been reached.
.T.G
Go: Instrument has been started.
.T.GC
GoCommand: Instrument has received a go command.
.T.S
Stop: Status 'Stop' has been reached.
.T.B
Begin of sequence.
.T.F
Final: End of determination, the final steps will be carried out.
.T.E
Error. Message together with error number, see page 60ff.
.T.H
Hold: Status 'Hold' has been reached.
.T.C
Continue: Continue after hold.
.T.O
Conditioning OK: EP reached (in SET with conditioning).
.T.N
Conditioning Not OK: EP not reached (in SET with conditioning).
.T.Re
Request: In the inquiry of an identification or the sample size after
start of titration.
.T.Si
SiloEmpty: Silo empty, i.e. the last line has been removed from
the silo memory.
.T.EP
EPList: Entry into EP list (with SET)
.T.RC
Results have been recalculated.
Messages for changings in the I/O lines. If the changings are made simulta-
neously, there is 1 message. Pulses receive 2 messages: one message each
for line active and inactive.
.I
Input: Change of an input line.
.O
Output: Change of an output line (except 7, pin 2, for recorder
pulses).
If a change occurs that requires a message, the Titrino sends space (ASCII
32) and ! as an introducer. This is followed by the name of the device (see
3.2.2.47). Special ASCII characters in the device name are ignored. If no de-
vice name has been entered, only ! is sent. Finally the Titrino sends the infor-
mation which node has triggered the message.
Example: !John".T.Si": The message was triggered from instrument "John",
node .T.Si
3.2.2.87. Setup.PowerOn
Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.
3.2.2.88. Setup.Initialise
Setup.Initialise.Select
Setting of default values for the following areas:
ActMeth:
Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Silo:
The silo memory is deleted. Same function as delete entire silo.
Calib:
pH calibration data for all measuring inputs.
Config:
All values under &Config.
Assembly: All values under &Assembly.
3.2 Remote control commands
ActMeth, Silo, Calib, Config,
Assembly, Setup, All
$G
$G
93