3.2. Remote control commands
82
Parameters for CAL: Calibration at Titrino.
With "ON", there are no hold points in the calibration sequence for entries, the
first buffer is measured directly.
3.2.2.36. Mode.Parameter.Calibration.ActPulse
Parameters for CAL: Output of a pulse on the I/O line "Activate", see page
131.
3.2.2.37. UserMeth.FreeMem
Memory space, available for user methods or silo lines. $Q sends the number
of free bytes, e.g. "4928".
3.2.2.38. UserMeth.Recall
UserMeth.Recall.Name
UserMeth.Store
UserMeth.Store.Name
UserMeth.Delete
UserMeth.Delete.Name
UserMeth.DelAll
Management of the internal method memory: Load, store and delete methods.
An action is performed if "$G" is sent to the corresponding node just after en-
tering the name.
Do not use blank characters before and after method name!
.DelAll: Deletes all methods in the user memory.
3.2.2.39. UserMeth.List.1.Name
UserMeth.List.1.Mode
UserMeth.List.1.Quantity
UserMeth.List.1.Bytes
UserMeth.List.1.Checksum
for each method
List of the methods in the user method memory with the following character-
istics:
.Name:
Name of the method
.Mode:
Mode
.Quantity: Measured quantity
.Bytes:
Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 3.2.2.59.
3.2.2.40. Config.PeriphUnit.CharSet
Selection of the character set and the graphics control characters of the Ti-
trino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.41. Config.PeriphUnit.Balance
Config.PeriphUnit.Plot
first, all, OFF
read only
up to 8 ASCII characters
up to 8 ASCII characters
up to 8 ASCII characters
read only
read only
read only
read only
read only
Epson, Seiko, Citizen, HP, IBM
Sartorius,Mettler,Mettler AT,
AND,Precisa
U, dU/dt, V, dV/dt, U(rel), T
719 S Titrino
$G
$G
$G
$G