Metrohm 719 S Titrino Instrucciones Para El Uso página 89

Tabla de contenido
719 S Titrino
.UnitSmpl: If "no unit" is desired for the sample size, the blank string must be
entered.
3.2.2.55. SmplData.ONSilo.DelLine
SmplData.ONSilo.DelLine.LineNum
Deletion of a silo line. The line # is deleted with &SmplData.ONSilo.DelLine
$G. If a formerly deleted line is edited again, it becomes valid (function "unde-
lete").
3.2.2.56. SmplData.ONSilo.DelAll
Deletes the entire silo memory. Must be triggered with $G.
3.2.2.57. SmplData.ONSilo.CycleLines
Silo data cycling.
With "ON", executed lines are copied to the next free silo lines, see page 53.
Exercise caution if you edit the silo memory during the determinations!
3.2.2.58. Info.Report
Info.Report.Select
$G sends the selected report to the COM which is set in
&Config.PeriphUnit.RepToComport:
configuration: Configuration report. Is not accessible during a running deter-
mination.
parameters: Parameter report of the current method. During a running deter-
mination only "live"-parameters are accessible.
smpl data: Current sample data.
statistics: Statistics table with the individual results.
silo:
Contents of the silo memory.
calib:
Calibration data of the measuring input in the current method.
C-fmla:
Contents of the <C-fmla> key.
def:
Contents of the <def> key.
user method: Contents of the method memory.
full:
Full result report of the last completed determination.
short:
Short result report of the last completed determination.
calc:
Calculation report of the current method.
all:
All reports.
ff:
Form feed on printer.
Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.59. Info.CalibrationData
Info.CalibrationData.Inp1.pHas
Info.CalibrationData.Inp1.Slope
Info.CalibrationData.Inp1.Temp
3.2 Remote control commands
configuration, parameters, smpl data,
statistics, silo, calib, C-fmla, def,
user method, full, short, calc, all, ff
-20.00...7.00...+20.00
-9.999...1.000...+9.999
-170.0...25.0...+500.0
$G
1...99, OFF
$G
ON, OFF
$G
$G
85