719 S Titrino
C44: (Last measured) temperature in °C. Used for the temperature
compensation in pH measurements.
C45: Start volume with SET in ml, "2.800".
C46: Asymmetry pH of CAL, "6.89".
C47: Relative electrode slope of CAL, "0.9950".
3.2.2.63. Info.StatisticsVal.ActN
Info.Statistics.1.Mean
Info.Statistics.1.Std
Info.Statistics.1.RelStd
etc. up to .9
The current values of the statistics calculation.
$Q sends, e.g.
ActN: Current value of the individual results
Data for MN1:
Mean: Mean value (decimal places as in result)
Std: Standard deviation (1 decimal place more than in result)
RelStd: Relative standard deviation (in %, 2 decimal places)
3.2.2.64. Info.ActualInfo.Inputs.Status
Info.ActualInfo.Inputs.Change
Info.ActualInfo.Inputs.Clear
Info.ActualInfo.Outputs.Status
Info.ActualInfo.Outputs.Change
Info.ActualInfo.Outputs.Clear
Status sends the current status of the I/O lines, Change sends the information
regarding whether a change in status of a line has taken place since the last
clearing, Clear clears the change information. For the output, there is a con-
version from binary to decimal, e.g.
| 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |
Line No.
| 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
Output: 2
1
+ 2
3
= "10"
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also pages 129ff):
Inputs:
0
Start (pin 21)
1
Stop (pin 9)
2
Enter (pin 22)
3
Clear (pin 10)
4
Smpl Ready (pin 23)
5
pin 11
6
pin 24
7
pin 12
3.2.2.65. Info.ActualInfo.Assembly.CyclNo
$Q sends the current cycle number of the voltage measurement cycle, e.g.
"127". From the cycle number and the cycle time (see 3.2.2.71), a time frame
can be set up.
The cycle number is set to 0 on switching on the instrument, on every start
and for QuickMeas. It is incremented as long as the instrument remains
switched on.
3.2 Remote control commands
Outputs:
0
Ready (pin 5)
1
Cond. ok (pin 18)
2
Titration (pin 4)
3
EOD (pin 17)
4
Freely selectable (pin 3)
5
Error (pin 16)
6
Activate (pin 1)
7
Pulse for recorder (pin 2)
read only
read only
read only
read only
"3"
"3.421"
"0.0231"
"0.14"
read only
read only
$G
read only
read only
$G
read only
87