3.2. Remote control commands
80
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.23. Mode.Parameter.Presel.Cond
Mode.Parameter.Presel.DriftDisp
Parameters for SET:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
3.2.2.24. Mode.Parameter.Presel.DCor.Type
Mode.Parameter.Presel. DCor.Value
Parameters for SET:
.Type:
Mode of drift correction
.Value:
drift value in µL/min.
If "auto" is set, the actual drift value at the start of titration is used for drift cor-
rection.
Formula for drift correction: EP – (drift value * titration time)
3.2.2.25. Mode.Parameter.Presel.IReq
Mode.Parameter.Presel.SReq
Parameters for SET, MEAS: Automatic inquiry after the start of the determina-
tion. From such an inquiry, the determination continues if the requested en-
try/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.73) or with &M
$G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.26. Mode.Parameter.Presel.ActPuls
Parameter for SET:
Output of a pulse on the I/O line "Activate", see page 131.
3.2.2.27. Mode.Parameter.Measuring.SignalDrift
Mode.Parameter.Measuring.UnitSigDrift
Mode.Parameter.Measuring.EquTime
Parameters for MEAS: Criteria for the measured value acquisition. Measured
value drift in mV/min (with pH, U, Ipol), µA/min (with Upol), resp. °C/min
(with T). Equilibration time in s. OFF means that the corresponding criterion is
switched off. If both criteria are OFF, the measurement continues indefinitely.
If the equilibration time has never been edited, it is automatically calculated by
the instrument to match the drift, see page 34. After it has been edited once, it
remains in force with the set value.
3.2.2.28. Mode.Parameter.Measuring.MeasInput
Mode.Parameter.Measuring.Ipol
Mode.Parameter.Measuring.Upol
auto, man, OFF
0.0, 99.9
id1, id1&2, all, OFF
value, unit, all, OFF
first, all, cond., OFF
pH, U, Ipol, T:
0.5...999, OFF
Upol:
0.05...99.9, OFF
read only
0...9999, OFF
1, 2, diff.
±127...1...+127
±1270...400...+1270
ON, OFF
ON, OFF
ON, OFF
719 S Titrino