3.2. Remote control commands
84
$G sets all RS settings. The changes are performed only if the instrument is
inactive. After the setting of the interface parameters, wait at least 2 s to allow
the components to equilibrate.
Settings of the values for the data transmission via the RS interface: baud
rate, data bit, stop bit, parity and type of handshake, see also page 95ff.
The setting of the values must be initiated with $G immediately after entry of
the values.
3.2.2.50. Config.ComVar.C30
with up to .C39, etc.
Values of the common variables from C30 up to C39. Insert the common
variables directly or describe the determination results directly from the
method, see 3.2.2.6.
3.2.2.51. SmplData.Status
On/off switching of silo memory. When the silo memory is switched on, the
sample data are fetched from the lowest valid silo line.
3.2.2.52. SmplData.OFFSilo.Id1
SmplData.OFFSilo.Id2
SmplData.OFFSilo.Id3
SmplData.OFFSilo.ValSmpl
SmplData.OFFSilo.UnitSmpl
Current sample data.
The identifications Id1...Id3 can be used in formulas as sample-specific calcu-
lation constants C21...C23.
If "no unit" is desired for the unit of the sample size, the blank string must be
entered.
3.2.2.53. SmplData.ONSilo.Counter.MaxLines
SmplData.ONSilo.Counter.FirstLine
SmplData.ONSilo.Counter.LastLine
Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.
3.2.2.54. SmplData.ONSilo.EditLine.1.Method
SmplData.ONSilo.EditLine.1.Id1
SmplData.ONSilo.EditLine.1.Id2
SmplData.ONSilo.EditLine.1.Id3
SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point
SmplData.ONSilo.EditLine.1.UnitSmpl
etc., up to .99
Contents of a silo line.
.Method: Method used to process the sample, from the method memory or
from the card.
.Id:
The identifications Id1...Id3 can also be used as sample-specific
calculation constants C21...C23 in formulas.
0... +999 999
ON, OFF
up to 8 ASCII characters
up to 8 ASCII characters
up to 8 ASCII characters
6-digits, sign and decimal point
up to 5 ASCII characters
read only
read only
read only
up to 8 ASCII characters
up to 8 ASCII characters
up to 8 ASCII characters
up to 8 ASCII characters
up to 5 ASCII characters
719 S Titrino