9 Cleaning and care

9
Cleaning and care
24
Operating Instructions SIROTorque L
Cleaning and disinfection of the
+
SIROTorque L
housing
The outer surfaces of the SIROTorque L
be cleaned and disinfected after each patient. If any
deposits and/or dirt residues are visible, you also must
clean the housing during treatment.
+
The SIROTorque L
can also be simultaneously cleaned
and disinfected with a disinfectant, e.g.
MinutenSpray classic from Alpro.
To clean and disinfect the SIROTorque L
moisten a clean, lint-free cloth in the disinfectant and use
it to wipe the outer surfaces of the housing, even in
hard-to-access areas. Check to make sure that all visible
dirt and residue has been removed from the surface. Let
the surfaces air-dry. Do not wipe them dry!
You can also use MinutenWipes from Alpro.
When cleaning and disinfecting the housing, make sure
that no liquids can penetrate into the interior of the unit.
Instead of using MinutenSpray classic or
MinutenWipes, you can also disinfect the
+
SIROTorque L
housing using a disinfectant
containing one of the following substances:
Alcohol, ethanol < 50%
Glutaraldehyde > 0.019%
ATTENTION
• Do not use disinfectants containing any of the following
substances:
– Phenol
– Alcohols > 50 %
– Peracetic acid
– Peroxide or any other oxygen-splitting agents
– Chlorine and iodine-splitting agents
• Check the gasket of the 4-hole highspeed handpiece
hose connection regularly and replace it if necessary.
• Regularly check all gaskets for damage or signs of wear
and tear, e.g. cuts, tears, notches or wear. Replace all
damaged or heavily worn gaskets as soon as they
appear to be worn out.
Sirona Dental Systems GmbH
+
+
housing must
+
housing,
61 74 499 D 3489
D 3489.201.02.02.02