ATTENTION
Please disconnect the power plug from the power supply
before performing cleaning or care of the unit!
Cleaning and disinfecting the water paths
+
The SIROTorque L
unit and the supply hose for the BL
or BL ISO electric motor are equipped with an internal
water supply which transports the spray from the dental
treatment center to the straight/contra-angle handpiece.
+
SIROTorque L
has no self-cleaning or self-disinfecting
system for this internal water supply. The internal spray
+
lines of the SIROTorque L
and the supply hose are,
however, connected to the spray supply of the dental
treatment center.
+
The SIROTorque L
spray lines should be cleaned and
disinfected together with the dental treatment center
once a week using Dentosept P (Order No. 33 18 156)
as described below:
+
Flush the SIROTorque L
least 20 seconds.
Then switch the treatment center and the
+
SIROTorque L
off for 24 hours.
+
Flush the SIROTorque L
30 seconds.
+
The SIROTorque L
and the BL or BL ISO motor must
remain connected to the spray supply of the treatment
center throughout the entire cleaning and disinfection
process.
i
NOTE
Use only instruments that are suitable for the specified
speed range!
ATTENTION
Be especially careful during endodontic procedures or
when working near the pulp.
For more information on the cleaning and care of electric
motors, please refer to the operating instructions
supplied with the relevant motor.
9 Cleaning and care
with Dentosept P for at
with water for at least
25