Troubleshooting - Sirona Dental SIROTorque L+ Manual Del Operador

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Sirona Dental Systems GmbH
Operating Instructions SIROTorque L
11

Troubleshooting

Error
Foot switch reacts too sensitively
Light on electric motor not lit
Instrument light is too dark or too bright
After prolonged operation (especially at
speeds < 15,000 rpm) the power decreases
61 74 499 D 3489
D 3489.201.02.02.02
+
See section 6.4 "Air pressure for the lower switching threshold on the
foot switch", increase air pressure.
a) Switch on instrument light, see page 17 "Operation – Switching the
instrument light OFF/ON"
b) Halogen lamp inside the electric motor is defective, install new halogen
lamp, see operating instructions of the corresponding electric
motor.
Adjust light intensity, see section 6.2 "Instrument light – lamp voltage
(light intensity)".
Internal protective circuit reduces current consumption to protect the motor
from overheating.
• Remove foot from foot switch for approx. 1 minute, allow motor to cool
down.
Full power is then restored.
Alternatively:
• Select a handpiece with reduction gear.
11 Troubleshooting
Remedy:
27