Cal Due Date
The Solaris Multigas Detector (with data logging and software version
1.1 or higher) is equipped with a Calibration Due Date feature. To
activate this feature, see Chapter 3, "Accessing the Instrument Setup
Mode".
If the calibration due date feature is activated, following Last Cal Date,
the message "CAL DUE, X DAYS" appears on the instrument LCD.
•
x = the number of days until a calibration is due,
user selectable for 1 to 180 days.
If the number of days until calibration is due reaches 0, an alert occurs
and "CAL DUE, --NOW--" displays.
•
Press the RESET button to clear the alert and continue with the
instrument warm-up period.
During Normal Measure mode, if the calibration due date feature is
activated and calibration is due, the instrument beeps and displays
"CAL DUE" every 30 seconds until the unit is calibrated.
Perform a calibration check before each day's use to verify proper
instrument operation (see Chapter 2, "Calibration Check").
Instrument Alarm Bypass Options
The Solaris Multigas Detector (with software version 1.1 or higher) is
equipped with a feature to disable or silence the visual, backlight,
audible, and vibrator options. To activate this feature, see Chapter 3,
"Accessing the Instrument Setup Mode".
If any of these options (visual, backlight, audible, or vibrator) are
disabled during instrument startup, the Solaris Detector displays:
•
"VISUAL OFF" if the red LEDs are disabled
•
"BACKLITE OFF" if backlight is disabled
•
"AUDIBLE OFF" if audible buzzer is disabled
•
"VIBRATE OFF" if the vibrator is disabled.
If the visual, audible, or vibrator options are disabled, "ALARM OFF"
flashes on the LCD during Normal Measure mode.
2-2