Turn-ON or
•
MIN reading was reset.
•
MIN appears in the upper portion of the display.
•
To reset the MIN Reading:
1. Access the Min page.
2. Press the RESET button.
Short Term Exposure Limits (STEL) (FIGURE 2-9)
Figure 2-9. Exposure Page with STEL Alarm
•
The STEL flag appear in the upper portion of the display to show
the average exposure over a 15-minute period.
•
When the amount of gas detected by the Solaris Multigas Detector
is greater than the STEL limit:
•
Alarm sounds
•
Alarm lights flash
•
STEL flashes.
To Reset the STEL:
1. Access the STEL page.
2. Press the RESET button.
The STEL alarm is calculated over a 15-minute exposure. Calculation
examples are as follows:
•
Assume the Detector has been running for at least 15 minutes:
2-14