Instructions - Stryker SofCare Manual De Operación Y Mantenimiento

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 91
1.
Place control unit on flat surface or suspend control unit on end of bed using attached hooks. See Figure 2 and
Figure 3. Remove the plug to disconnect the device.
2. Position the mattress on bed frame.
3. Connect the hose assembly between the mattress air cell and the control unit. Unscrew the cap from the air valve
of the mattress and screw the adaptor from control unit onto the air valve tightly. The mattress could work either
with or without the check valve.
4. Plug the power cord and adjust the pressure control knob to highest setting for quick inflation and turn control unit
on using green on/off switch. The unit will take approximately 40 minutes to inflate the mattress.
5. After installation, make sure the flap is not folding upwards to avoid fluid seeping through mattress cover.
NOTE
Make sure the control unit is suitable for the local power voltage and frequency.
6. Position patient on the mattress and adjust the pressure control knob for patient comfort.
7.
A Hand Check must be performed every 8 hours to verify proper operation of the device. See Figure 6.
8. To perform the Hand Check:
With the patient on his or her back, slide hand, flat and palm up, between the overlay and the mattress. Hand should
be directly under the air cell that is under the patient's buttocks (or other bony area). See Figure 6.
Wait for full inflation of the air cell directly above hand. If the patient's body is not in direct contact with hand, the system
is operating correctly. If, during full inflation of air cell, the patient's body is in direct contact with flat hand, the system is
not operating properly. Adjust the pressure control to a higher setting. Wait 10 minutes and repeat the Hand Check. If
the Hand Check fails, check that the hoses are not kinked or pinched. If repeated Hand Check fails and hoses are not
kinked, contact Stryker for further instruction.
WARNING
Deflate before CPR or CPR could be ineffective.
To deflate mattress for CPR:
Unscrew the cap screw and pull the tab to remove the check valve from the air valve if any. The air cell will deflate in
approximately 20 seconds. Proceed with CPR procedures.
www.stryker.com

Instructions

▼ Figure 6
HAND CHECK
Properly inflated
Under-inflated
▼ Figure 7
476005-5210 V1.6
Return To Table of Contents
11
Tabla de contenido
loading

Este manual también es adecuado para:

2890

Tabla de contenido