2. Insert the pin into the nose of the collet or into the
back cap of the handpiece until the desired length
pin extends from the nose.
NOTE: Make sure the pin extends far enough through
the collet to allow the collet to hold the pin before
squeezing the clamping lever.
3. Remove the handpiece from the SAFE mode to
allow rotary drill operation.
4. Place the handpiece trigger in the FORWARD
orientation as required.
5. Squeeze the clamping lever against the handpiece
handle to grasp pin.
6. Hold the pin against the bone. Press the
FORWARD handpiece trigger to drive the pin into
the bone.
7. To obtain additional pin length perform the
following steps.
7. 1. Release the clamping lever of the collet.
7. 2. Pull the handpiece back from the bone to
allow more pin length between the handpiece
and bone.
7. 3. Squeeze the clamping lever to grasp the pin.
7. 4. Press the FORWARD handpiece trigger to
drive the pin into the bone.
www.stryker.com
Print Date: Sep 21, 2016 02:12:36 PM
0000000958, Rev. F Effective Date: Sep 21, 2016 11:14:29 AM
To Remove the Pin
1. Place the handpiece trigger in the REVERSE
orientation as required.
2. Squeeze the clamping lever to grasp the pin.
3. Pulling the handpiece and pin away from the bone,
press the REVERSE handpiece trigger to remove
the pin from the bone.
To Remove the Attachment
See the instructions for use supplied with the
handpiece.

Care Instructions

For processing instructions and disposal/recycle
information, see the care instructions manual supplied
with the handpiece.
Troubleshooting
See the instructions for use supplied with the
handpiece.