Man" And "Timer" Mode Faulty Cases - Amprobe MultiTest 2000 Manual De Usuario

Ocultar thumbs Ver también para MultiTest 2000:
Tabla de contenido
AMPROBE

6.2.4. "MAN" and "TIMER" mode faulty cases

If
the
instrument
detects the External
Power supply adapter
the
instrument
show the message
displayed to the right.
If
the
instrument
detects
a
Voltage
between
the
terminals higher than
15V, the instrument
does not perform the
test and displays the
screen to the right for
5 seconds.
MΩ
will
-.- -
---V
ν REMOVE POWER
MAN
500V
FUNC
VNOM
MΩ
input
-.- -
---V
ν VOLT IN INPUT
MAN
500V
FUNC VNOM
This result can't be saved
05.06.01
Disconnect
M
Power Supply Adapter
15s
05.06.01
ATTENTION: the test can't be
executed. Check that the circuit is
M
not energized.
15s
MULTITEST2000