APPENDIX 1 – MESSAGES DISPLAYED
Message
AUTONOM:
CLEAR ALL? (Enter)
CLEAR LAST? (Enter)
Data saved
DATA SIZE:
HOLD
Password:
Invalid date
Energy Measuring
Memory Full
No ext supply!
No parameter sel
No Phase selected
PASSWORD ERROR
PASSWORD OK
Please wait
Recording
Too many param
Description
Available memory autonomy for the recording which is
being effected
The operator is trying to cancel all the recordings effected
The operator is trying to cancel the last recording effected
The data have been saved
Dimensions of the stored data
By pressing the proper key, the HOLD function has been
activated
A recording has been started and at least 5 minutes have
passed from the last activity of the instrument (see
paragraph 7).
The inserted date is not correct
The instrument is taking an energy measurement
The memory of the instrument is full
A recording has been started without connecting the
external power supply (optional code A0050)
A recording has been started without selecting any value to
be recorded
Voltage and/or current harmonics have been selected and
the corresponding flag has been enabled (HARMONICS
ON) but no phase voltage or current has been selected
The inserted password is wrong (see paragraph 7).
The inserted password is correct
The instrument is waiting for the recording to be started
(see paragraph 6)
The instrument is recording (see paragraph 6)
More than 63 parameters have been selected (harmonics
included) or More than 38 parameters with CO-
GENERATION Flag enabled
MULTITEST2000
Advices
Press ESC in order not to cancel the whole
memory, press ENTER to confirm
Press ESC in order not to cancel the last
recording, press ENTER to confirm
Press HOLD again to disable this function
Insert the password:
F1, F4, F3, F2
Check the inserted date
Press F1 to stop it
Cancel some recordings after transferring
them to a PC
Verify it you really want to start the recording
without the external power supply. In that
case press START again.
Press START/STOP and select at least a
value entering the MENU
Select at least one phase voltage and/or
current
Check the password
Deselect some values