Measurement Procedure - Amprobe MultiTest 2000 Manual De Usuario

Ocultar thumbs Ver también para MultiTest 2000:
Tabla de contenido
AMPROBE
6.1.2.

Measurement Procedure

1. Select the desired mode using the F1 key.
2. Connect the black and blue test leads to B1 and B4 input terminals respectively
Connection of the test leads during LOWΩ test.
3. If the cables supplied with the instrument are not long enough for the measurement
you can extend the blue cable.
4. Short-circuit the test leads making sure that the conductive parts of the alligator clips
make a good contact to each other. Press the START key. If the display doesn't
show 0.00Ω repeat the test leads calibration (see paragraph 6.1.1).
5. Connect the instrument terminals to the ends of the conductor under test (see
previous picture).
6. If the mode "RT+" or "RT-" was selected use the F3, F4 keys to set the
duration of the test.
7. Press the START key. The instrument will execute the measurement. In RT+/RT-
START
(Timer mode) you can press START key again if you want to stop the test before the
STOP
duration set is expired.
ν
WARNING:
Never
disconnect
"Measuring" is displayed.