Measurement Procedure - Amprobe MultiTest 2000 Manual De Usuario

Ocultar thumbs Ver también para MultiTest 2000:
Tabla de contenido
AMPROBE
6.2.
MΩ:
INSULATION RESISTANCE MEASUREMENT WITH 50V, 100V, 250V,
500V, 1000V TEST VOLTAGE
The measurements comply with IEC 61557-2 and VDE 0413 part 1.
ν
ATTENTION:
Turn the rotary knob to the MΩ position.
The F1 key allows the operator to select one of the following measuring
modes:
MAN mode (Manual mode) Recommended test.
TMR mode (Timer mode: test duration depends on the selected interval
from 10 to 999 seconds). This test can be executed when the test
required a defined duration.

6.2.1. Measurement Procedure

1. Select the desired mode using the F1 key.
2. Connect the test leads to the instrument input terminals B1 and B4 respectively,
Before performing an insulation test make sure that the circuit
under test is not energised and all the loads are disconnected.
MULTITEST2000

Capítulos

Tabla de contenido
loading

Tabla de contenido