Maintenance
4
Maintenance
4.1
Recommended test and maintenance intervals
The following information complies with DGUV-R 112-190 valid in
Germany. Observe national directives.
Type of work to be
performed
Check by the device user
Cleaning and disinfection
Visual inspection, function
and leak test
Replace exhalation valve
disc
Replace speech diaphragm
Replace O-ring of the speech
diaphragm
1)
For full face masks in air-tight packaging, otherwise every 6 months
2)
For full face masks in air-tight packaging: every 2 years
18
X
1)
X
X
2)
X
X
X
X
(approved by Dekra Exam)
4.2
Cleaning and disinfection
Clean and disinfect the full face mask after every use.
WARNING
!
The mask body must be cleaned sufficiently thoroughly that no
residues of skin care products remain on the mask body.
Otherwise the elasticity in the sealing area can be lost and the
facepiece will no longer sit correctly.
CAUTION
!
Do not use solvents (e.g. acetone, alcohol) or cleaning agents
containing abrasive particles for cleaning and disinfection. Use
only the described procedures and the specified cleaning
agents and disinfectants. Other agents, doses and contact
times may cause damage to the product.
For
information
disinfectants and their specifications, refer to document
9100081 at www.draeger.com/IFU.
X
1. Remove accessories and clean separately.
2. Prepare a cleaning solution consisting of water and a cleaning
X
agent.
3. Clean all parts using a soft cloth with cleaning solution.
4. Thoroughly rinse all parts under running water.
5. Prepare a disinfectant bath consisting of water and a disinfectant.
6. Put all parts to be disinfected in the disinfectant bath.
7. Thoroughly rinse all parts under running water.
8. Dry all parts in the air or in a drying cabinet (temperature: max.
60 °C). Do not expose to direct sunlight.
9. Fit the accessories.
about
suitable
cleaning
agents
Dräger X-plore 6000