• Provide the correct voltage to the system through the power cord.
• The system is equipped with a main switch (thermal breaker) able to protect the system components
against overload or short-circuits. Move the breaker switch to the position "1" to start operating with the
TPS-mobile.
• The system activation by means of the Main Switch provides voltage to TMP controller but both TMP and
Foreline pump continue to be switched off.
The system activation allows the TMP controller to start reading the pressure data through the optional
FRG-700 Full Range Gauge if it is connected to the gauge connection port on the rear side of the system.
• It is possible to start pumping by means of the selector named "Pumping". As soon as the Pumping se-
lector is moved to the position "1" the TMP and Foreline pump will be switched on.
PNs 9698404, 9698405, 9698406, 9698419, 9698430, 9698431, 9698432, 9698433, are not equipped with
the selector named "Pumping", the same function is provided by the pushbutton Start/Stop on the TMP con-
troller.
SYSTEM ACTIVATION
(with thermal protection)
GAUGE CONNECTION
START PUMPING
"Pumping" selector
Note:
A1-2/3
Main switch
Gauge port
Serial port
APPENDIX 1
87-901-005-01(A)