Error codes

Caution
The error code updates without announcement.
Error
Meaning
code
E00
No clock
parameter
E03
Pressure zero
error
E04
Low battery
E05
Inflation error
E06
Above 320mmHg
E07
Controlled stop
using STOP key
E08
Pulsation can not
be measured
E10
Pulsations can not
be detected
because the
patient may have
moved.
E20
Pulse rate < 30
200 < Pulse rate
E21
DIA < 40
160 < DIA
E22
SYS < 60
280 < SYS
DSD < 10
E23
150 < DSD
Error Code
Status
All parameters are lost.
Reset status.
An error code is displayed
without cuff inflation.
Measurement is stopped.
An error code is displayed.
Auto mode is quit.
Inflation pressure does not
reach the target pressure.
An error code is displayed.
Air is exhausted. An error
code is displayed.
Measurable pulsation is
searched to 20mmHg in
constant exhaust. An error
code is displayed.
In the measurement, Quick
exhaust is executed. An
error code is displayed.
An error code is displayed.
DIA
: Diastolic Blood Pressure
SYS
: Systolic Blood Pressure
DSD
: The Difference between Systolic Blood Pressure
:
and Diastolic Blood Pressure.
Page 32
Operation and Treatment
Enter clock parameters. Refer to
"Setup of display and clock"
Release the air from the cuff
completely.
Replace with new batteries.
Restart the auto mode if you use it.
Wrap the cuff and connect to main
unit exactly. If you can not clear the
error, there may be an air leak and
repair is necessary
Do not move and try to relax during
the measurement. If you can not
clear the error, the product will
requier repair.
Do not press the STOP key if you do
not need to use it.
Do not move and try to relax during
the measurement. The error occurs
when measurable pulsations are not
received due to thick cloth or quick
motion.
Do not move and try to relax during
the measurement.
Measure the blood pressure by other
methods.