14.3 Fault on the unit
Fault
Unit does not switch on
Unit switches off again af-
ter a short time
Loud operating noises af-
ter switching on lasting for
more than 30 seconds
Unit not responding
Unit is on, but there is
no display on the touch

screen

2151100018L40 1903V012
OBJ_DOKU-0000003578-mehrere-011.indd 53
Probable cause
No mains voltage
On/off switch is defective
Mains cable or power supply unit
plug not inserted correctly
Hardware defect
Mains supply voltage too low
Radiation deflector defective
The unit has not yet completed
the startup procedure
Unit is blocked by the firewall
Touch screen initialization fault
Touch screen brightness set too
dark
Touch screen defective
Troubleshooting
Solution
Check the mains cable and
i
plug connection and replace if
necessary
Check the power supply unit
i
If the touch screen does not
i
light up, replace the power
supply
Check the mains fuse in the
i
building
Contact technician
i
Check the mains cable and plug
i
connections
Contact technician
i
Check the mains supply voltage
i
Contact technician
i
After switching the unit on, wait
i
20 - 30 seconds for the startup
procedure to be completed
Enable the ports for the unit in
i
the firewall settings
Switch the unit off and on again
i
Update firmware
i
Increase the brightness of the
i
touch screen
Contact technician
i
09.04.2019 10:51:22
EN
US
53