11.4 Scanning image data via the

touch screen on the unit

Starting the imaging plate scanner
No PC connection is needed for scanning the
image data via the touch screen The image
data is stored locally on the memory card In or-
der to transfer the image data to the imaging
software, the unit must be connected to a com-
puter
There are two options for scanning via the touch
screen:
Scan:
Before scanning the image data, the
patient data and exposure settings of
the image are acquired and then saved
with the image data
If no patient data and exposure set-
tings of the image are entered, the im-
age is saved to a folder with date and
time
Quick scan:
The image data is saved to a folder
with the date and time and no addi-
tional information
Use Help on the touch screen for further
information on operating the unit via the
touch screen
Prerequisites:
– Memory card (SDHC, max 32 GB) is in-
serted in the slot of the unit
Press
to switch the unit on
i
2151100018L40 1903V012
OBJ_DOKU-0000003578-mehrere-011.indd 39
Start scanning:
On the touch screen, tap on Scan
i
Enter the patient data
i
Select the image settings and scanning mode
i
An animation on the touch screen prompts
for insertion of the imaging plate
Insert the imaging plate only once the
bar above the animation is green
Figure 8: Example of the animation
prompting for insertion of the imaging plate
Start quick scan:
On the touch screen, tap on Quick scan
i
Select the scanning mode
i
An animation on the touch screen prompts
for insertion of the imaging plate
Insert the imaging plate only once the
bar above the animation is green
Figure 9: Example of the animation
prompting for insertion of the imaging plate
Usage
EN
US
39
09.04.2019 10:51:19