Usage
Scanning an intraoral phosphor storage plate
CAUTION
Loss of image data caused by stray
ambient light entering the unit
Remove the plate guide only after the
i
phosphor storage plate has dropped
EN
into the collecting tray
US
Slide the phosphor storage plate into an un-
i
occupied plate guide until the PSP is auto-
matically drawn in
The inactive side must be visible
40
OBJ_DOKU-0000003578-mehrere-011.indd 40
Scanning progress is displayed on the touch
screen Image data is saved automatically
The touch screen only shows a pre-
view as an initial impression of the X-
ray image Limitations to image pre-
views occur due to image size and/or
exposure conditions For purposes of
diagnosis, the X-ray image must be
viewed on a diagnostic monitor
After it has been scanned, the phosphor
storage plate is erased and drops into the
collecting tray
Save the X-ray image
i
Remove the PSP and prepare it for recording
i
another X-ray image
Remove the plate guide after the entry slot
i
status LED on the touch screen switches to
green
Reading the extraoral image plate
In order not to mix up X-ray images, only
scan the X-ray images from the selected
patient
Remove the plate protector from the X-ray
i
cartridge
Place the plate protector with the phosphor
i
storage plate in the entry slot of the transport
drum
2151100018L40 1903V012
09.04.2019 10:51:19