Place the PSP in the barrier envelope into the
i
patient's mouth
Doing this, make sure that the active side of
the PSP faces the X-ray tube
Set the exposure time and setting values on
i
the X-ray unit (see "83 X-ray unit settings")
Record the X-ray image
i
Result:
The image data must be scanned within 30 min-
utes
Preparing for scanning
CAUTION
Light erases the image data on the
phosphor storage plate
Never handle exposed phosphor stor-
i
age plates without either a barrier en-
velope or an intraoral/extraoral plate
protector
Wear hand protection
Remove the PSP with the barrier envelope
i
from the patient's mouth
WARNING
Contamination of the unit
Clean and disinfect the barrier enve-
i
lope before removing the PSP
In the event of heavy soiling, e g from blood,
i
dry clean the barrier envelope and protective
gloves, e g wipe with a clean cellulose cloth
2151100018L40 1903V012
OBJ_DOKU-0000003578-mehrere-011.indd 35
2
e
s it e s id
O p p o
.
s tu b e
to w a rd
2 1 4 2
8 7
4 7 5
D 4 8
a
Disinfect the barrier envelope and protective
i
gloves with a disinfection wipe
Alternatively, use a spray disinfectant on a
soft, lint-free cloth
Allow the barrier envelope to fully dry
i
Pull off the protective gloves, disinfect and
i
clean the hands
NOTICE
Powder from the protective gloves
on the PSP damages the unit during
scanning
Completely clean all traces of the pro-
i
tective glove powder from your hands
before handling the PSP
Tear the barrier envelope and place the phos-
i
phor storage plate in the transfer box
If the active side of the phosphor storage
plate is facing up, turn the phosphor storage
plate over immediately