5. Commissioning and operation
5.3 Tightening torques
Only ever screw in, or unscrew, the instrument via the spanner flats and to the
prescribed torque using an appropriate tool.
The correct torque depends on the dimensions of the connection thread and the seal
used (form/material).
NPT (tapered) threads are marked with "NPT" on the thread. Parallel threads do not
have any marking.
Screwing or unscrewing the connection head is not permitted.
When screwing in the instrument, please observe that the threads are not skewed.
If M20 x 1.5 threads are loosened to the head with a counter nut, the IP degree of
protection can no longer be guaranteed.
NPT threads to the head must not be loosened.
A PTFE/PFA lubricant is applied at the NPT threads to the thermowell/protection
tube. This must be renewed when the thread has been loosened.
Number of threads engaged according to the respective certification.
WIKA operating instructions, models TR10-L, TC10-L, TR10-0, TC10-0 (Ex d)
Connection head, (1/4000, 7/8000, PIH)
Cable gland (metal)
Thread
1/2 NPT
3/4 NPT
M20 x 1.5
The connection between connection head and neck
tube must not be disconnected by the user.
The connection head may only be aligned via the
nipple-union-nipple threaded connection.
As-delivered condition: Hand-tightened (approx. 5 Nm)
After aligning the connection head, the connection
must be secured using 50 ... 60 Nm
Thread
1/2 NPT
3/4 NPT
G 1/2 B
G 3/4 B
M14 x 1.5
M18 x 1.5
M20 x 1.5
M27 x 2
Tightening torques
20 Nm
20 Nm
20 Nm
Tightening torques
30 Nm
40 Nm
35 Nm
40 Nm
27 Nm
35 Nm
37 Nm
41 Nm