Soxtec™ 2050 Automatic System
6.5.3 Replacement of the Condensers
1.
Follow step 1-4 in "Replacement of the front glass".
2.
Lift the front glass and put it beside the instrument.
3.
See Fig. 6:3. Disconnect the cooling tubings (pos 1) and the teflon drain tubings
(pos 2). These drain tubings must also be changed when changing condensers.
4.
Unscrew the condenser holder (pos 3).
5.
Unscrew the air nipple (pos 4).
6.
Unscrew the magnet holder (pos 5) and the plastic distance (pos 6). After this
unscrew the screw (pos 7) holding the rod (pos 8).
7.
Disconnect the condenser, start with the nut (pos 9) and take away the holder
(pos 10) and the spring (pos 11).
8.
Lift the rod and change the condensers (pos 12), one by one.
9.
Assemble in reverse order.
Note: Always use the same sample rod for respective condenser
to avoid leakage.
6.5.4 Replacement of Seals
See section 4.3.2 Installation / Change of Seals on page 4:3.
Note: Replacement of electrical components and wires should
be performed by a service engineer authorized by FOSS Analyt-
ical AB.
In need of service contact FOSS Analytical AB or your local representative.
User Manual 1000 7414 / Rev. 4
6:5