PAR. 14
SETTING THE DATE AND TIME
To set the and time which are visualised on the display, proceed as follows:
1) Turn on the autoclave with the ON/OFF main switch, keeping pressed the 0/1 button. After some seconds
on the TIME/MESS. display, the writing –1- is visualised comparing the first programming phase. Now you
can release the 0/1 button.
On the TEMPERATURE display, three digit are visualised. The first digit not blinking show this parameters: 0
seconds – 1 minutes – 2 hours – 3 day – 4 month – 5 year. The two next blinking digit show the values to
selected. Press the 12A button to change the selection; press the 11A button to increase the value, or the
14A button to decrease it. Once the wished value has been reached, press the 0/1 button, to pass at the
following programming.
2) The digit –2- is visualised. On this page, it is possible to set the atmospheric air alignment. On the
PRESSURE display, the writing 0.0 blinks. Keep the door opened and press the 0/1 button, to confirm the
datum. This is needed to level the atmospheric pressure which changes according to altitude.
3) Press the 0/1 button, to confirm the data and pass to the following programming page; on the display the
digit –3- is visualised. On this page, it is possible to set the type of external units connected to the system
(TEMPERATURE display). Pressing 11A button or the 14A button to increase or respectively decrease the
number of connected units: 0=None – 1=Internal Printer – 2=External Printer – 3=Computer – 4=Internal
Printer+Computer. Setting DEFAULT = 2.
4) Press the 0/1 button, to confirm the data and pass to the following programming page. On this page, the
serial number of the autoclave is set. Do not change this setting.
Press the 0/1 for five seconds to exit the programming.
27
DPTA208 Rev_2