Emissions
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Advanced is intended for use in the electromagnetic
environment specified below. The customer or the user of the breast pump should assure that it is used in
such an environment. The essential performance functions are the accuracy of the vacuum pressure and the
backflow protection to withstand milk leakage into the motor unit. The Philips Avent single/double electric
breast pump Advanced was tested for immunity to electromagnetic disturbances and passed using the
criteria of no change in modes, vacuum accuracy (±20 %), backflow protection and operation of the breast
pump.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
RATED power frequency
magnetic fields IEC
61000-4-8
Electrical fast transient/
burst
IEC 61000-4-4
Surges
IEC 61000-4-5
Compliance
Class B
Complies
Class A Category
Complies
Immunity test level
±8 kV contact
±15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
30 A/m
50 Hz or 60 Hz
±2 kV
100 kHz repetition
frequency
±1 kV Line-to-line
±2 kV Line-to-ground
Electromagnetic environment –
guidance
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power
supply network which supplies buildings used
for domestic purposes.
Electromagnetic environment -
guidance
Floors can be a source of static electricity. When
the air is very dry (<20 % RH) the user is advised
to prefer to use the device in a room with
concrete or wooden floor (instead of using it in
a room with a synthetic floor).
The device is immune against any RF wireless
communication device such as smart phone, Wi-
Fi or Bluetooth device at a distance of 30 cm
(11.8 in) or more. Avoid a closer distance as this
may result in improper operation.
Mains power quality expected of a typical
commercial or residential environment.
Mains power quality expected of a typical
commercial or residential environment.