Tecno-gaz ANDROMEDA VACUUM xp Instrucciones De Uso página 32

Idiomas disponibles
  • ES

Idiomas disponibles

ENGLISH
09
CYCLES
STERILIZATION
C1
Not wrapped
134°C
C2
Wrapped
134°C
C3
Not wrapped
121°C
C4 Wrapped
121°C
C5
SPECIAL
134°C
* The values indicate are the maximum load for every tray, for ferrous materials or solid or
hollow part
** The values indicate are the maximum load for every tray, for porous parts
It is compulsory to wrap the devices aimed at an invasive use (for instance; Surgical
instruments, etc.) to ensure a higher sterility.
The data of this table are indicative; the choice of the sterilisation cycle must be
based on the data supplied by the producer of the object you want to sterilise.
10
TURN ON THE DEVICE PRESSING THE MAIN SWITCH.
Connect the suitable supplied hose to the loading plug and insert the other end (the one
with the filter) inside the demineralised or distilled water tank.
The use of feedwater with contaminants at levels exceeding those given
in the table above can greatly shorten the working life of the device, with
NOTE:
damage at the components of the autoclave and then invalidate the
manufacturer's warranty.
Pressing the button "16A", the pump is automatically turned on. By sucking, the pump fills
the tank.
The pump supplies water for 120 seconds; after reaching the maximum level, it is turned
off.
In case during these 120 seconds the pump does not reach the maximum level, the
operator must press the button "16A" up to switching off the pump in automatic mode (for
the maximum reached level).
The autoclave does not work in case the tank level in on MIN. as signalled by the LED.
The attempt to start the autoclave with MIN. level state is visualised with the message
"ER2" on the display "TIME/MESS".
10
STERILIZATION TABLE
TIME OF
TIME OF
DRYING
6
17
11
17
18
17
21
17
6
2
FILLING OF WATER TANK
CHARGE*
CHARGE**
(Kg)
(g)
1.25
50
1.25
50
1.25
50
1.25
50
1.25
50
WORK PRESS.
(bar, differential
pressure)
2,1÷2,2
2,1÷2,2
1,1÷1,2
1,1÷1,2
2,1÷2,2