10.1 Characterisstics of the water to be used
TABLE OF QUALITY LEVEL IN ACCORDING TO DIRECTIVE DIN EN 285
Evaporation residue
Silicium oxide, SiO2
Iron
Cadmium
Lead
Rest of heavy metals except iron, cadmium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20°C)
PH value (degree of acidity)
Appearance
Hardness (E Ions of alkalin earth)
11
In case the pump does not work, the loading tank can be filled by the operator in the
following way:
Turn on the device.
Unscrew the cap.
Insert the supplied pipe for manual filling outside of which you can find a funnel.
Pour the distilled water in the funnel which is kept higher thank the loading cap;
Pour up to when the pilot light "LEVEL WATER TANK, max." is turned on.
Screw the cap checking its tightening
12
The device is fitted with a tank for collecting the waste waters used during the sterilisation
cycle; this tank is separated from the loading tank.
When the discharge tank is full, the max level pilot light "EXHAUST TANK", is turned on.
To discharge the tank, proceed as follows:
Connect the supplied pipe to the discharge tap.
Unscrew the discharge tap in counterclockwise direction.
Flow the waste waters.
Screw the discharge tap without forcing it and checking it is closed properly.
In case of sterilization failure, the used water in discharge tank, can have contamined
residual. This water must de disposal in the black water, use protettive latex glover to
carry out the operation.
The device does not work in case the discharge tank is full. In case the operator tries to
start a cycle even in this state, the device locks visualising "ER3" on the "TIME/MESS".
display.
CEN STANDARD DIN EN 285
MANUAL FILLING OF THE WATER TANK
EMPTYING THE TANK
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5 to 7
Colourless clean without sediment
0.02
ENGLISH
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
s/cm
mmol/l
11
loading